A method for counting monosodium urate crystals in synovial fluid
Abstract
This study was aimed to standardize the technique for counting monosodium urate (MSU) crystals in the synovial fluid (SF) of patients with gout. A total of 52 SF specimens were examined under a polarized light microscope. The amount of SF ranged between 0.1 and 45 mL (median 3 mL). MSU crystals were counted in four areas with the same size at 400x magnification. Cytological examination of the same specimens was also performed. Median leukocyte count was 400 cells/mm3 (range 50-14,000 cells/mm3), with a median percentage of polymorphonuclear leukocytes of 9% (range 0%-98%). Median crystal count was 179.5 (range 3-1600). Inter- reader and intra-reader agreement in crystal counting were good with a weighed k of 0.89 [95% confidence interval (CI) 0.85-0.94] and 0.89 (95% CI 0.84-0.93), respectively. Our data indicate that the SF MSU crystal count is a feasible and highly reliable technique.Downloads
Download data is not yet available.

Published
2015-06-30
Keywords:
Gout, Monosodium urate crystals, Synovial fluid, Polarized microscopy.
Statistics
- Abstract views: 1421
- PDF: 644
How to Cite
Montagna, P., Brizzolara, R., Ferrone, C., Cutolo, M., Paolino, S., & Cimmino, M. (2015). A method for counting monosodium urate crystals in synovial fluid. Reumatismo, 67(1), 29-32. https://doi.org/10.4081/reumatismo.2015.827
Copyright (c) 2015 P. Montagna, R. Brizzolara, C. Ferrone, M. Cutolo, S. Paolino, M.A. Cimmino